beautypg.com

Technical data, General technical data valid for all pumps, Gas inlet temperatures – VACUUBRAND MZ 2S NT User Manual

Page 25: Conditions (see also “technical data”, pg. 25), Protection, “technical data”, pg. 25), Substances, see section “technical data”, pg. 25

background image

page 25 of 91

Technical data

General technical data valid for all pumps

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

Maximum permissible installation altitude

above mean sea level

ft

(m)

6500

(2000)

No-load speed 50/60 Hz

rpm

1500 / 1800

Device fuse

slow blow fuse 6.3 A

Motor protection

thermal cutout, manual reset

Degree of protection IEC 529

IP 40

A-weighted emission sound pressure

level* (uncertainty K

pA

: 3 dB(A))

dB(A)

45

Gas inlet temperatures

We reserve the right for technical modification without prior notice!

* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-

mate vacuum with silencer at outlet.

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

This manual is related to the following products: