Notice – VACUUBRAND MZ 2S NT User Manual
Page 58
Service kit for ME 2 NT................................................................................696877
Service kit for ME 4 NT / MZ 2 NT ..............................................................696860
Service kit for MD 4 NT / MV 2 NT .............................................................696861
Service kit for ME 8 NT .............................................................................. 696862
Service kit for MZ 2D NT ............................................................................696863
Service kit for ME 4S NT / MZ 2S NT ..........................................................696868
ME 8S NT / MD 4S NT ...............................................................................2x 696868
Service kit for ME 4R NT ............................................................................696859
Service kit for MD 4CRL NT ....................................................................... 696870
Diaphragm key (width 66 mm) ....................................................................636554
+
Allow sufficient cooling of the pump before
starting maintenance.
+
Vent the pump and isolate it from the
vacuum system before you start mainte-
nance.
Ensure that maintenance is done only by suitably
trained and supervised technicians.
The valves and diaphragms as well as the motor capac-
itors are wear parts. If the rated ultimate vacuum is no
longer achieved or in case of increased noise level, the
pump interior, the diaphragms and the valves must be
cleaned and the diaphragms and valves must be checked
for cracks or other damage.
All bearings are encapsulated and are filled with long-life
lubricant. Under normal operating conditions, the drive
system is maintenance free.
In demanding circumstances, it may be efficient to check
and clean the pump heads on a regular basis. In normal
use, the lifetime of the diaphragms and valves is typically
15,000 operating hours.
- Prevent internal condensation, transfer of liquids or
dust. The diaphragms and valves will be damaged if
liquid is pumped in significant amount.
- Carry out maintenance frequently if the pump is ex-
posed to corrosive media or in case of deposits.
- Regular maintenance will improve the lifetime of the
pump and also protect both users and the environment.
NOTICE
atm