Operating conditions, Notice – VACUUBRAND MZ 2C NT +AK+M+D User Manual
Page 15
page 15 of 74
+
Do not use this product in an area where it can fall or
be pulled into water or other liquids.
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, installation in higher altitudes, conductive
pollution or external condensation on the pump.
• Do not operate this product near flames.
To the best of our knowledge the equipment is in com-
pliance with the requirements of the applicable EC-direc-
tives and harmonized standards (see ”Declaration of Con-
formity”) with regard to design, type and model. Directive
EN 61010-1 gives in detail the conditions under which the
equipment can be operated safely (see also IP degree of
protection, ”Technical data”, pg. 24).
Operating conditions
➨
These pumps are not approved for operation in po-
tentially explosive atmospheres.
Do not operate the
pumps in potentially explosive atmospheres.
➨
Pumps without the
”
`
”
mark on the rating plate are
not approved for the pumping of potentially explo-
sive atmospheres. Do not pump potentially explo-
sive atmospheres with those pumps.
➨
Pumps bearing the
”
`
”
mark on their rating plates
are
approved for the pumping of potentially ex-
plosive atmospheres according to their classifica-
tion
II 3G IIC T3 X according to ATEX, but they are
not approved for operation in potentially explosive
atmospheres(see section „
Equipment marking (ATEX)“, pg. 22)
.
➨
The pumps are
not suitable to pump any of the sub-
stances listed below.
NOTICE