Troubleshooting, Fault possible cause remedy – VACUUBRAND BVC basic G User Manual
Page 29

page 29 of 35
Fault
Possible cause
Remedy
❑ No aspiration.
➨ Filter clogged?
✔ Replace filter.
➨ Leak at the bottle head? ✔ Check seals, coupling
and blind cap. Screw in
or replace if necessary.
➨ No vacuum supply con�
nected?
✔ Connect vacuum supply.
➨ No vacuum while vacu�
um supply connected?
✔ Check vacuum supply.
Open lock valve in the
vacuum line if applica�
ble.
❑ Pressure control
device switches fre�
quently.
➨ Leak in the system?
✔ Check tubing, seals,
coupling, and blind cap.
Screw in or replace if nec�
essary.
➨ Filter clogged?
✔ Replace filter.
➨ Vacuum controller con�
taminated inside or
defective?
✔ Contact local distributor.
Troubleshooting
This manual is related to the following products: