beautypg.com

Maintenance and repair, Notice – VACUUBRAND BVC basic G User Manual

Page 10

background image

page 10 of 35

• Use the system only in combination with the integrat-

Use the system only in combination with the integrat�

Use the system only in combination with the integrat�

ed hydrophobic filter (sterilizable) to protect equipment

and environment (people).

Maximum liquid level in collecting bottle: approx. 80%,

depending on the application. (When working with low-

boiling liquids, or liquids that tend to foam, the maximum

liquid level might be reduced).

Maintenance and repair

Attention: The equipment might be contaminated with

noxious or otherwise dangerous process chemicals

that have been pumped during operation. Ensure that

the equipment is decontaminated before maintenance.

+

Take adequate precautions to protect people from the

effects of dangerous substances if contamination has

occurred. Wear appropriate safety-clothing when you

come in contact with contaminated components.

+

Never operate a defective or damaged device.

In order to comply with law (occupational, health and safe�

ty regulations, safety at work law and regulations for en�

vironmental protection) vacuum pumps, components and

measuring instruments returned to the manufacturer can

be repaired only when certain procedures (see section

”Notes on return to the factory”) are followed.

NOTICE

NOTICE

This manual is related to the following products: