beautypg.com

VACUUBRAND BVC 21 NT VARIO User Manual

Page 22

background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999157 / 10/12/2009

page 22 of 31

View of the disassembled pump head parts

To check valves use hex key to remove four socket head

screws from pump head and remove upper housing

(housing cover with housing cover insert), head cover and

valves.

Position the pumping unit onto its left side with the pump

head at the top to facilitate maintenance.

Never remove parts by using a spiky or sharp-edged tool

(e.g. screw driver), we recommend to use a rubber mallet

or compressed air (to be blown carefully into port).

Remove head cover from housing cover insert and check

valves. Note position of valves and remove.

Replace valves if necessary.

Check diaphragm for damage and replace if necessary.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.

9

1

2

4

5

7

6

8

3

1: Housing

2: Connecting rod

3: Washer

4: Diaphragm support disc

5: Diaphragm

6: Diaphragm clamping disc with

square head screw

7: Head cover

8: Valves

9: Housing cover with housing

cover insert

This manual is related to the following products: