User’’s Guide | TS-700 MS

19

Error Codes

ERROR CAUSE

SOLUTION

uuuuuu

1. Overloaded condition.
2. Wrong connection at

load cell.

3. Load cell has quality

problem.

1. Reduce the weight.
2. Check load cell

connection

3. Inspect load cell.

Check the input
and output.

nnnnnn

1. Calibration is incorrect.
2. Bad connections.
3. Load cell has quality

problem

.

1. Check scale is

resisted or not, foot
is kept level or not.

2. Check load cell

connection.

3. Check input and

output resistance of
load cell

.

ERR1

Overweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.

1. Input weight

correctly during
calibration.

2. Place correct

weight on weighing
platform.

ERR2

Underweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.

The calibration
weights Minimum is
10% of Max. cap.
Recommend 60%-
80% of Max. Cap.

ERR3

During calibration, the
input signal is negative.

Check all
connections.
Check load cell.
If connections and
load cell are okay,
PCB needs
replacing.

ERR4

During calibration, the
signal is unstable.

After the platform is
stable, start
calibration.

ERR5

Circuitry error

Replace the PCB.