Error Codes
ERROR CAUSE
SOLUTION
uuuuuu
1. Overloaded condition.
2. Wrong connection at
load cell.
3. Load cell has quality
problem.
1. Reduce the weight.
2. Check load cell
connection
3. Inspect load cell.
Check the input
and output.
nnnnnn
1. Calibration is incorrect.
2. Bad connections.
3. Load cell has quality
problem
.
1. Check scale is
resisted or not, foot
is kept level or not.
2. Check load cell
connection.
3. Check input and
output resistance of
load cell
.
ERR1
Overweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.
1. Input weight
correctly during
calibration.
2. Place correct
weight on weighing
platform.
ERR2
Underweight calibration
error.
Incorrect weight input or
incorrect weight on
weighing platform.
The calibration
weights Minimum is
10% of Max. cap.
Recommend 60%-
80% of Max. Cap.
ERR3
During calibration, the
input signal is negative.
Check all
connections.
Check load cell.
If connections and
load cell are okay,
PCB needs
replacing.
ERR4
During calibration, the
signal is unstable.
After the platform is
stable, start
calibration.
ERR5
Circuitry error