Triner Scale TS 700-MS Series 2 User Manual
Page 14

User’s Guide | TS-700 MS
12
C20
CAUTION:
Changing
these
parameters will
affect the
indicator’s
performance.
Zero range
(amount that
can be zeroed
out with the
Zero key)
0 = disable manual zero
setting
1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity
C21
Zero range at
power-up
0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
C22
CAUTION:
Changing
these
parameters will
effect the
indicator’s
performance.
Automatic Zero:
tracking range
0 = disable zero tracking
0.5 = ±0.5d
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note:
• d = division
• The zero tracking range
cannot be more than
manual zero range.
C23
Automatic
Zero; tracking
time
0 = disable zero tracking
time
1 = 1 second
2 = 2 seconds
3 = 3 seconds
C24
CAUTION:
Changing
these
parameters will
effect the
indicator’s
performance.
Overload range
00 = close overload range
01d -- 99d
(d = division)