User’’s Guide | TS-700 MS

12

C20

CAUTION:
Changing
these
parameters will
affect the
indicator’’s
performance.

Zero range
(amount that

can be zeroed
out with the
Zero key)


0 = disable manual zero

setting

1 = ±1% max capacity
2 = ±2% max capacity
4 = ±4% max capacity
10 = ±10% max capacity
20 = ±20% max capacity
100 = ±100% max capacity

C21

Zero range at

power-up

0 = no initial zero setting
1 = ±1% max capacity
2 = ±2% max capacity
5 = ±5% max capacity
10 = ±10% max capacity
20 = ±20% max capacity

C22
CAUTION:

Changing
these
parameters will
effect the
indicator’’s
performance.


Automatic Zero:

tracking range








0 = disable zero tracking
0.5 = ±0.5d
1.0 = ±1.0d
2.0 = ±2.0d
3.0 = ±3.0d
4.0 = ±4.0d
5.0 = ±5.0d
Note:

•• d = division
•• The zero tracking range

cannot be more than
manual zero range.

C23

Automatic

Zero; tracking
time

0 = disable zero tracking

time

1 = 1 second
2 = 2 seconds
3 = 3 seconds

C24
CAUTION:

Changing
these
parameters will
effect the
indicator’’s
performance.

Overload range

00 = close overload range

01d -- 99d

(d = division)