Shellab SVAC2E-2 User Manual
Page 10

10
OPERATION
7.1
Power Supply: Connect the service cord to a grounded outlet and push the power 
switch to the ON position. 
7.2
Place a reference thermometer inside the chamber where it can be easily viewed 
through the window, then vacuum down the chamber as described in Vacuum Operation 
Section. 
7.3
Setting Temperature: The operating range for this oven is 40
C - 210
C. To set the
Temperature Controller turn the control knob to the desired oven temperature, using the 
graduated dial as a reference guide. Allow at least one hour for the temperature to 
stabilize. Using a reference thermometer, verify the oven temperature, if it is not the 
correct value, turn the control knob up or down as needed. Allow the temperature to re-
stabilize, and continue this process until the exact desired temperature is achieved. 
NOTE: Slight vapor or smoke may occur in the initial heat-up. This is a normal 
occurrence when the oven is first brought up to temperature and protective coatings on 
the element heat up. 