Display failure, Operation (continued), Isplay – Shellab SSI2-2 User Manual
Page 12: Ailure

Page | 12
OPERATION (CONTINUED)
High Speed Error
This “rErrH” message denotes a high speed error. Typically, excessive speeds happens when most
or all of the containers are removed from the shaker while it is in operation.
Solution: If a container has been removed, replace the container or restart the machine.
Electrical Error
The
“Error” message indicates an electrical error. This occurs if the unit has lost power during a
run.
Solution: Check that the power cord is properly installed, check that the power supply circuit has
not been interrupted, and restart.
D
ISPLAY
F
AILURE
If the display will not illuminate, check the following:
Electrical connection - disconnect main power cord and connect again.
Voltage supply - check the voltage.
Verify voltage on the data plate on the back of the incubator.
Blown fuse - check the fuse.
Defective power switch - check the voltage at switch.