10

5.13

Operating Lights Manually Without Running Program: To turn lights on without running a program
ENT1 (event output 1) must be turned on. Follow instruction below on how to turn the output on.

20.0

20.0

PRESS AND

HOLD BOTH UP

AND DOWN

ARROWS FOR

3 SECONDS

OR UNTIL

NEXT SCREEN

OPer

A

USE UP

OR DOWN

ARROWS

TO SELECT

NEXT SCREEN

DISPLAY

OPer

P.StA

PUSH

ADVANCE KEY

REPEATEDLY

UNTIL NEXT

SCREEN
DISPLAY

APPEARS

oFF

Ent 1

USE UP

OR DOWN

ARROWS

TO SELECT

on IN TOP

DISPLAY

on

Ent 1

PRESS INFINTE

KEY REPEATEDLY

TO RETURN

TO HOME

PAGE

20.0

20.0

To turn the lights off follow the same procedure but, turn ENT1 to off.

PID SETTINGS (from factory)

P.I.D.

h.Pb

12.0

c.hy

0.2

ti

180

td

50

db

0.2

=
=
=
=
=

The PID values or stored in the loop menu under operation page. These parameters control how the unit
will control at set point. The incubator is setup from the factory to cycle the compressor on and off to keep
the evaporator from freezing up to eliminate damage to the coil from off gassing from Fruit Fly growth. Due
to the cycling of the compressor the temperature will fluctuate up and down above and below set point. To
run the incubator at a stable temperature, change the parameter c.hy to -5.0. This will make the compressor
to run continuously and should make the incubator run stable at set point. All other parameters are control
sensitive and should not be changed. To enter the loop menu follow the instruction under calibration
procedure.

5.14

Set Up Parameters: The setup parameters are parameters that configure the control on how it functions
and should not be changed. Below is the list of parameters that is sent from the factory for reference. They
should never be changed unless there is a complete understanding of what they do and how they will affect
control functions. For better understanding on what these functions do refer to the Watlow manual on the
Watlow disk.

diO SET

LOOP SET 1

FUN 1

FUN 2

diO 5

diO 6

n.Ag

pid

LEV

HIGH

LEV

HIGH

dir

otPt

dir

otPt

Ca.g

ON Off

fn

P.Str

Fn

NONE

FN OFF

FN

OFF

h.Pb

12.0

Fi

I

Fi

0

c.hy

0.2

A1 SET

t1

180

SEN

RO.1H

td

50

gLbL SET

Rtl

2

db

02

C.F

C