Pid settings (from factory) p.i.d, P.sta – Shellab SRI21FV-2 User Manual
Page 10

10
5.13
Operating Lights Manually Without Running Program: To turn lights on without running a program
ENT1 (event output 1) must be turned on. Follow instruction below on how to turn the output on.
20.0
20.0
PRESS AND
HOLD BOTH UP
AND DOWN
ARROWS FOR
3 SECONDS
OR UNTIL
NEXT SCREEN
OPer
A
USE UP
OR DOWN
ARROWS
TO SELECT
NEXT SCREEN
DISPLAY
OPer
P.StA
PUSH
ADVANCE KEY
REPEATEDLY
UNTIL NEXT
SCREEN
DISPLAY
APPEARS
oFF
Ent 1
USE UP
OR DOWN
ARROWS
TO SELECT
on IN TOP
DISPLAY
on
Ent 1
PRESS INFINTE
KEY REPEATEDLY
TO RETURN
TO HOME
PAGE
20.0
20.0
To turn the lights off follow the same procedure but, turn ENT1 to off.
PID SETTINGS (from factory)
P.I.D.
h.Pb
12.0
c.hy
0.2
ti
180
td
50
db
0.2
=
=
=
=
=
The PID values or stored in the loop menu under operation page. These parameters control how the unit
will control at set point. The incubator is setup from the factory to cycle the compressor on and off to keep
the evaporator from freezing up to eliminate damage to the coil from off gassing from Fruit Fly growth. Due
to the cycling of the compressor the temperature will fluctuate up and down above and below set point. To
run the incubator at a stable temperature, change the parameter c.hy to -5.0. This will make the compressor
to run continuously and should make the incubator run stable at set point. All other parameters are control
sensitive and should not be changed. To enter the loop menu follow the instruction under calibration
procedure.
5.14
Set Up Parameters: The setup parameters are parameters that configure the control on how it functions
and should not be changed. Below is the list of parameters that is sent from the factory for reference. They
should never be changed unless there is a complete understanding of what they do and how they will affect
control functions. For better understanding on what these functions do refer to the Watlow manual on the
Watlow disk.
diO SET
LOOP SET 1
FUN 1
FUN 2
diO 5
diO 6
n.Ag
pid
LEV
HIGH
LEV
HIGH
dir
otPt
dir
otPt
Ca.g
ON Off
fn
P.Str
Fn
NONE
FN OFF
FN
OFF
h.Pb
12.0
Fi
I
Fi
0
c.hy
0.2
A1 SET
t1
180
SEN
RO.1H
td
50
gLbL SET
Rtl
2
db
02
C.F
C