LABEC M-PL-600 User Manual
Page 18

Appendix 1 Checking method of pH linear error
Input 177.48mV(pH4.00 buffer solution electrode potential simulate
value)、-178.07mV(pH10.01 buffer solution electrode potential simulate
value) into the instrument,temperature compensation sent to 25℃,
operate according to 2 points calibration;after calibration the instrument
displays “
”signal.(if type 640 please set the buffer solution set to
b1)Check the linear error of the instrument according to following table.
mV
414.11 354.95 295.80 236.64 177.48 118.32
59.16
0
pH
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
mV
-414.11 -354.95 -295.80 -236.64 -177.48 -118.32 -59.16
pH
8.00
9.00
10.00
11.00
12.00
13.00
14.00
- 17 -
See also other documents in the category LABEC Equipment:
- AA13 (7 pages)
- GW1060 (44 pages)
- SS1 (4 pages)
- AA-8BS (30 pages)
- AA-22BS (32 pages)
- AA-22N (32 pages)
- AA-8S (33 pages)
- AA-22S (26 pages)
- Shinva-45B (21 pages)
- AA9 (6 pages)
- USC-5L (5 pages)
- WS1.0 (3 pages)
- BEM4 (4 pages)
- LHF-40 (5 pages)
- NBCT2 (4 pages)
- NBCT7S (4 pages)
- ICB-12 (3 pages)
- LBC-4 (4 pages)
- SB-SDC-6 (5 pages)
- J-NBT (16 pages)
- BOB1 (5 pages)
- M-SOD-3636 (3 pages)
- J-SWB35 (21 pages)
- FXP16M (13 pages)
- M-SI-100 (3 pages)
- M-TMO1600 (3 pages)
- M-HOG-020 (13 pages)
- M-HOG010 (12 pages)
- J-USRC (16 pages)
- Z-L-450 (17 pages)
- Z-L-450A (6 pages)
- Z-L-530 (20 pages)
- M-SCEN-206 (13 pages)
- M-SCEN-207 (12 pages)
- Z-ZHWY-2102 (9 pages)
- GLF-90 (2 pages)
- PLCF300 (2 pages)
- H-DW-HW50 (12 pages)
- DW-FL90 (8 pages)
- B-IM0070FSC (5 pages)
- GLR-70 (3 pages)
- PLRGD-1400PT (3 pages)
- PVR-130 (3 pages)
- SPR-100 (3 pages)
- CEMLS (2 pages)