LABEC M-PL-600 User Manual
Page 16

good quality, then recalibrate the electrode. If the error still occurs, 
please change the electrode and calibrate the electrode. 
Err9:During 2 point/3point calibration, instrument judge 1point ( or 2 points)
which used buffer solution are same ( or the electrode has 
damaged), the instrument will judge the reliable calibration 
operation and modify the relative electrode parameter. If this 
occurs, please recalibrate the pH electrode. 
 
11. Maintenance of the instrument 
The performance of the instrument, is not only rely on the self
construction of the instrument, but also on good maintenance. Especially 
for pH meter, which high impedance is needed, and corrosive chemical 
agents is sometimes contaminated, therefore, necessary maintenance is a 
must 
Please, only when the instrument is power off, the plug of the adaptor
could be removed.
Please keep the instrument away from directly sunshine, otherwise
the life time of the LCD will be shortened.
The pH value of the sample solution should be between the pH buffer
solutions you selected. If the temperature of the buffer solution during 
calibration is same as the temperature of the sample, the highest 
resolution will be reached by the instrument. 
Temperature compensation function in the instrument is only 
effected to compensate the sample temperature changes 
according to pH sensor. 
The electrode plug should always keep high cleaning and dry. In order
to protect the electrode from broken, please do not contact the head of the 
electrode with hard materials, since it is very thin. Please refer to the 
instruction manual of the electrode for the storage, cleaning and 
maintenance of the electrode. . 
 Please do not contact the central part of the BNC electrode plug by 
- 15 -