4. Temperature Calibration________________________
The Labec Warming Tray has been calibrated as a set. Warming
Tray can be calibrated by placing a calibrated thermometer on
the plate for 1 hour and then measuring the difference. Altering
the “PV0fS” value to match is then carried out. (see the
Appendix A manual for Calibrating and only perform this if more
than 2ºc difference is recorded)
5. Troubleshooting Guide__________________________
Many operating problems may be solved by reading and carefully
following the instructions in this manual. Some suggestions for
troubleshooting are given below. Should these suggestions not
resolve the problem, contact the SERVICE DEPARTMENT on
02-9560 2811 for assistance or email to [email protected]. Please
supply the serial number and include a full description of the
problem.
Problem Comments______
LED do not light up 1. Check the RCD
2. Ensure that the AC
power switch is ON.
3. Check the three-pronged
power cord is properly
plugged into a grounded
three-prong AC outlet
of the appropriate
voltage.
Laboratory Equipment Pty Ltd - Warming Tray 2014 Ver2.01 Instruction Manual
5