Pre-installation – LABEC H-DW-HW50 User Manual
Page 6

2.Pre-Installation
How to transport:Pick up the unit from the bottom of tank and a slant plane within 45° is
acceptable. All transportation should be carefully.
Remove all wrappers.
Check accessories and documents according to package list.
Clean the unit thoroughly before any operation.
Installing location:Do not position the unit in direct sunlight or near heating diffusers,
radiators, or other sources of heat. The ambient temperature should not exceed 32ºC. A
minimum 20cm space on the sides, rear and top should be provided for better air
circulation and heat discharge. Location where is too cold and place with high humidity or
can be easily splashed by water are forbidden.
On the flat floor, Can directly impulse the freezer moving at stone's throw.
Notice:When moving the freezer, Please notice the wheel not to damage the line.
Notice:Should take away the packing base before run the freezer.
Notice:When connect the power supply,Don’t put the goods in the freezer
immediately.Please waiting for 12 hours
The freezer working condition:
A
. The environment temperature:10ºC~30ºC
B
. The relative humidity:≤80%
C
. No strong shake and causticity gas around.
D
. No sunshine and other cold/hot power effect.
☞
Ultra Low temperature Freezer working rule
:
Running intermission
P3:N and E port can’t connect
l
Safety Introduction
①
Power supply:This ULT freezer is designed operating under 220V/50HZ AC power