beautypg.com

Grant Instruments FH16-D User Manual

Page 10

background image

Flow Heaters

Part No: 15630/Issue 3

Page

March 2003

10

6 Technical

Specification

This equipment is for indoor use and will meet its performance figures within an ambient
temperature range 10 to 35

°C with maximum relative humidity of 80%.

Installation category II (transient voltages). Pollution degree 2 in accordance with IEC 664. For
operation at altitudes up to 2000 metres.

FH16 - D

Temperature Range

Ambient 15°C to 80

°C (FC1 is required for operation

at or below ambient)


Stability (DIN 58966) at 37

°C

± 0.004


Supply voltage range

220-240V 50/60 Hz

110-120V 50/60 Hz

Power rating - 230V

0.85kW

Heater power - 230V

0.75kW

Overall dimensions l/w/h (mm)

340/205/140


Overtemperature protection is by variable resettable cut-out.


Pump Performance:

Max head 2.2m

Max flow 19 lit/min at zero head.

Pump bore 6mm

7

Maintenance and service


All Grant Laboratory products are designed to comply with IEC1010-1 and can be flash tested.
Some are fitted with radio frequency interference suppressers. Therefore it is recommended that
only a d.c. test is performed.

Inspect and replace where necessary any worn or damaged hoses.

The overtemperature cut-out should be checked periodically by turning the set overtemperature
control anticlockwise until the alarm lamp comes on. The cut-out should then reset and set-up. If
the alarm lamp fails to light with the knob fully turned anticlockwise the unit should be checked by
a competent person.

No other routine maintenance is required.

DO NOT RETURN THE TANK

7.1

Cleaning

The cases can be cleaned with a damp cloth after disconnection. Do not use solvents. The
immersed parts can be cleaned using proprietary heating element cleaners. CAUTION: these
may be toxic - follow the cleaner manufacturer's instructions.

Before using any decontamination or cleaning method except that recommended, check with our
Service Department, or in other countries with our distributor, that the proposed method will not
damage the equipment.