Fault diagnosis, Symptom possible cause action required – Grant Instruments BTDL User Manual

Page 10

background image

SYMPTOM

POSSIBLE CAUSE

ACTION REQUIRED

NO DISPLAY

Unit not switched on.

Unit not connected to power supply.

Power supply failure.

Fuse blown in unit or plug (UK units
only).

Switch on.

Connect to power supply.

Check that other electrical appliances
on the same circuit are working.

Check and replace (see 6.2).

5. Fault diagnosis

TEMPERATURE DOES

NOT RISE WHEN

EXPECTED

TEMPERATURE CONTINUES

TO RISE WHEN

NOT EXPECTED

Set temperature is lower than block
temperature.

Set temperature is too close to
ambient.

Thermal fuse has operated.

Temperature control circuit fault.

Check set temperature.

Raise set temperature.

Have BTD thermal fuse replaced by a
competent person.

Have unit checked by competent
person.

Set temperature is higher than the
block temperature.

Temperature control circuit fault.

Check setting.

Have unit checked by competent
person.

Page 10

BTD/BTDL

Operating instructions

Version 7 - May 2011