V. calibration procedures – Galaxy Audio CHECK MATE CM-130 User Manual

Page 8

background image

SOUND LEVEL METER

6

V. Calibration Procedures

Using a standard Acoustic Calibrator
(94dB, 1KHz sine wave)

70 dB
90 dB

130 dB

110 dB

100

60
80

40

LEVEL

SLOW

FAST

A / C

MAX

(1) Make the following switch settings.

Display: dBA
Time Weighting: FAST
Measurement Mode: MAX Mode function disable.
Level Range: 80 to 110dB

(2) Insert the microphone housing carefully into the insertion

hole of the calibrator.

(3) Turn on the calibrator and adjust the CAL potentiometer

of the meter until the level display indicates the desired
level. The meter is calibrated before shipment. We
recommend recalibration annually.

Screwdriver

Acoustic Calibrator