Appendix A: 

Calibration mode: 

 

Warning:  Only  trained  qualified  personnel  may  open  covers  and  performs  this 

procedure.  

 

To  calibrate  the  U70F  a  precision,  balanced  output,  amplitude  jitter‐free  sine‐wave 

enerator,  a  small  Philips  screwdriver,  and  a  very  small  flat‐head  screwdriver  are 

g

necessary. 

 

Step 1: Turn‐off the power switch located on the back‐panel. Disconnect the mains cord. 

small Philips driver. 

 

Step 2: Open the top‐cover after removing all screws with a 

Step 3: Engage all four switches on the front‐panel (labeled PEAK, HOLD, RANGE, and BAR)

Step 4: Re‐connect the mains cable and turn power back on 

Step 5: Disengage the two switches in the middle (labeled HOLD and RANGE) during the 

self‐test  and  calibration  cycle.  The  U70F  will  enter  calibration  mode  after  the 

completion of the self‐test. 

Step 6: Apply the desired balanced input signal (1 kHz sine‐wave), 1.23V for +4dBu or any 

other level between +2dBu and +20dBu. The display will either show all yellow 

LEDs (‐39 dB to +20 dB] illuminated if the level is too low, all green LEDs [‐9 dB to 

0 dB] if the level is correct, or all red LEDs [+1 dB to +10 dB] if the level is too high. 

Step 7: Adjust the Level controls located on the meter PCBs counter‐clockwise if the level 

is too low or clockwise if the level is too high. The level adjustment control takes 

10 turn from maximum to minimum, so please be patient and keep turning. The 

.1% (+‐1dB is about +‐

procedure allows a calibration accuracy of less than a +‐0

10%).  

tep 8: Press any button to return to the normal operating mode.  

tep 9: Turn the instrument off, disconnect power and put the top‐cover back on. 

S

S