Installation – Envirco Isoclean UV User Manual
Page 7

Page 3
H
OSPI
-G
ARD
®
I
SO
C
LEAN
®
and
I
SO
C
LEAN
®
UV
Operation Manual
INSTALLATION
2.0. Installation
2.1.
Unpacking Instructions (crated for international)
2.1.1. The Hospi-Gard lsoClean is shipped in an open-framed crate. The module
should be inspected for exterior shipping damage immediately upon arrival. If
any damage is observed, a claims report should be completed and promptly filed
with the responsible carrier.
2.1.2. Uncrate and remove all packing material. Examine the unit for internal damage.
lf damage is discovered inside the crating, file a claim with the responsible carrier
immediately. The shipping components list and actual material received should
be compared and any shortages reported to Envirco immediately.
2.2. Assembly
2.2.1. The Hospi-Gard lsoClean is shipped completely assembled from the factory and
requires no assembly. Every unit is tested prior to shipment to verify airflow,
noise level and that the unit is leak-free.
2.2.2. Place the unit in its final operating position (see Figure 1 in section 3.4. for
recommended positioning).
2.2.3. Lock the wheel locks on the casters to keep the unit from accidental re-
positioning.
2.3. Initial
Start-Up
NOTE: The CDC requires that quantitative leakage and filter performance tests be
performed at the initial installation and every time the filter is changed or
moved. Tests should be repeated every six months. The factory tests should
not be substituted for the initial test.