Operation Manual

Page 15

 

 

 

 

4.15. Unit

Decontamination

 

 

NOTE: Use proper protocol, gowning and protective measures during unit

decontamination.

 

Should decontamination of the unit be required, the following is an adaptation by
Annex G. of the Recommended Microbiological Decontamination Procedure from NSF
49, June 2008. Confirm with appropriate Biosafety and lndustrial Safety Professionals
that the procedures meet your facility's guidelines.

 

 

4.15.1. Paraformaldehyde Decontamination

 

4.15.1.1. Relocate the unit as described in Section 4.14 to a controlled

access, non-public area with a non-porous floor, good ventilation
and a dedicated exhaust directly outside the building. A new
prefilter is not required. Attach and seal a flexible hose to the
dedicated exhaust. Place the other end of the exhaust hose near
the unit.

 

4.15.1.2. Calculate the total volume of the unit by multiplying the height, width

and depth. The total lsoClean volume is 13 cubic feet (61 in. x 24
in. x 15.5 in.) or (1.55m x 0.61m x 0.39m).

 

4.15.1.3. Multiply the total volume of the unit by 0.3g/ft

3

to determine the gram

weight of Paraformaldehyde required. Decontamination of the
lsoClean requires 3.9 grams (12 ft

3

x 0.3g/ ft

3

) of

Paraformaldehyde.

 

4.15.1.4. Remove the intake grille and wipe it down with an appropriate

surface decontaminant.

 

4.15.1.5. Place an unplugged heating device, such as a commercially

available electric frying pan with the thermostat set at 232.2° to
246.1° C (450° to 475° F) inside the unit through the removed
prefilter access pane

l

. Spread the Paraformaldehyde evenly over

the heating surface of the electric frying pan.

CAUTION: The auto-ignition temperature of Paraformaldehyde is 300° C

o

(572 F). Place a hot plate, beaker of water, and temperature

and humidity indicators in the unit next to the pan that will
contain the Paraformaldehyde.

 

4.15.1.6. Enclose all sides of the unit with heavy gauge plastic film and tape in

place while leaving the film that will close the space formerly
occupied by the intake grille free. The power cord should be coiled
and taped to the unit and sealed under the plastic film. Seal the film
to the floor on which the unit stands.

 

4.15.1.7. Determine the temperature and humidity inside the unit.