Operation Manual

Page 11

 

 

 

 

CAUTION:

Proper protective equipment and measures must be used at all times

during unit cleaning and maintenance. Check with your Safety Office to
assure the cleaning solutions, protective equipment (disposable hospital
gown, HEPA respirator, protective gloves) and protocols followed comply
with your facility’s and CDC guidelines.

 

CAUTION:

Disconnect the unit from the electrical power source before attempting
service. Do not work inside the IsoClean

UV with the UV light on. Proper

replacement of UV containment baffles, after maintenance, is mandatory to
maintain safe short-term and long-term UV exposure levels.

 

4.2. Cleaning

the

Unit

 

4.2.1. Periodic Cleaning of the unit may be done with a mild disinfecting soap and water

solution. A 1% or 1/100 dilution of bleach is an optional disinfectant

4

.

 

4.2.2. Exterior cabinet cleaning must be performed with the unit running. Wipe the

exterior of the unit with a damp sponge or cloth, paying particular attention to
the prefilter intake grille.

 

 

4.3. Prefilter

Replacement

 

4.3.1. The prefilter should be replaced every 60 to 90 days. Regularly replacing the

prefilter will maximize the life of the HEPA filter.

The lsoClean UV

comes standard with an anti-microbial pleated prefilter. lt is

recommended that a pleated prefilter is used.

 

4.3.2. Wipe the unit as described in Section 4.2.

THE FOLLOWING STEPS MUST BE PERFORMED WITH THE UNIT (FAN) RUNNING:

 

4.3.3. Spray a mist of phenolic disinfectant such as "Amply " or "Lysol" into the air

intake grille.

 

4.3.4. Remove the two (2) thumb nuts from the face of the intake grille and pull

the grille out.

 

4.3.5. lnsert gloved hand into a red plastic, medical waste disposal bag. Grasp the

top portion of the prefilter with bagged hand and gently pull outward. The airflow
of the unit will offer a resistance to pulling the prefilter out.

 

4.3.6. Turn the bag inside out over the prefilter then seal the bag for containment.

 

4.3.7. lnstall a new prefilter and secure the grille by following step 4.3.4 in reverse order
of removal.

 

 
 
 

 

 
 

 

4

Rutala WA, APlC guideline for selection and use of disinfectants, Am J lnfect Control 1996 Aug ; 24(4): 313-42.