3 operational information – Buffalo Filter ViroVac User Manual
Page 5

In addition, please compare the accessories you receive with the standard accessories list below. If
an item is missing, please notify BUFFALO FILTER Customer Service.
Standard Accessories:
• Operator’s Manual
• Power Cord
• Pneumatic Footswitch
Please contact BUFFALO FILTER Customer Service to purchase the following accessories:
• Replacement Filters
• EZ Link Remote Activation Device
• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories
1.3 Operational Information
internationally:
100/120 VAC
50/60 Hz, 220/240 VAC 50/60 Hz
2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I
3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part
4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1
5. Method of sterilization or disinfection recommended by BUFFALO FILTER (UL 60601-1,
Clause 5.4):
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution
or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable
7. Mode of operation (UL 60601-1, Clause 5.6): Continuous
8. Upon request, BUFFALO FILTER will provide the following:
Service and Repair Instructions, including Circuit Diagrams and Parts List
cian as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo), (F1, F2)
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo), (F1, F2)