N2400306, August 2011, Rev 4 Shaking Hot Tub Operators Manual

7

3.3.9 The unit is programmed with an adjustable high temperature alarm. The

high temperature alarm value is factory defaulted at 5°C above the set
temperature value. To change this value, follow the steps below.

3.3.10 Depress the “F” switch four times, and the display will change to “ALAr”

and will alternate between “Alar” and the current high temperature alarm
value.

3.3.11 Depress the “S” switch and the display will now flash the current high

temperature alarm value.

3.3.12 If a different value is desired use the Increment / Decrement Dial to

obtain the desired high temperature alarm value.

3.3.13 When the desired alarm value is indicated, press “S” switch. This new

value is now stored.

3.3.14 If a high temperature alarm condition is encountered the display will flash

“ALAr”, the heater will shut off, and the unit will beep.

3.3.15 This condition will persist until the bath temperature drops below the high

temperature alarm set value, or the user raises the High Temperature
Alarm setting above the bath temperature. This can be accomplished by
depressing the “F” key; the display will alternate between “ALAr” and the
actual high temperature alarm value. Depress the “S” switch and the
display will flash the actual high temperature alarm value. Use the
Increment / Decrement Dial to obtain the new desired high temperature
alarm value, and depress the “S” key to store this new value.

Note: If a change to the temperature set point is made or the unit power is cycled
on and off, the high temperature alarm value will return to a default of 5°C above
the set temperature value.


3.5. Catastrophic Heater Condition

3.5.1 In the event that a chamber temperature would reach 120°C the systems

heater will be disabled and remain in the disable mode until the over
temperature condition is corrected.

3.6

Draining the Tank

3.6.1 Allow the liquid to cool to below 60°C.
3.6.2 Place the drain hose assemble into the drain port located on the rear of the

unit. As soon as the drain hose assemble is snapped into the drain port
the bath will begin to drain.

3.6.3 When the bath is drained, be sure to remove the drain hose assembly by

depressing button on drain port and remove hose from drain port.