Technical specifications, Maintenance and service, 1 cleaning – Boekel Scientific 133000S Economy Digital Incubators User Manual
Page 7: 2 replacement of fuses
![background image](/manuals/661606/7/background.png)
6
7. Technical Specifications
This equipment is intended for indoor use and will meet its performance figures within
the ambient temperature range of 10
°C to 35°C, with maximum relative humidity of 80%
(non-condensing). Installation Category II (transient voltages). Pollution Degree 2 in
accordance with IEC 664. Suitable for operation at altitudes of up to 6500 feet.
Specifications:
Temperature Range:
(Ambient +5
°C) to 60°C
Temperature Uniformity:
+/- 1
°C at 37°C
Temperature Display Resolution: 0.1
°C
Supply Voltage Range:
120V +/- 10%, 50/60 Hz
230V +/- 10%, 50/60 Hz
Power Rating:
Model 133000: 90W
Model 133000S: 90W
Model 133000-2: 90W
8. Maintenance and Service
All Boekel laboratory products are designed to comply with IEC1010-1. No routine
maintenance is required. There are no user serviceable parts in this product.
There are no parts that are required to be examined or supplied by Boekel or our
agent.
8.1
Cleaning
Disengage power cord prior to cleaning. If a spill occurs, use appropriate clean
up procedures as required for radiation or biohazard control. The outer casing
may be cleaned with water and a damp cloth. Do not submerge or immerse the
Incubator in water. Before using any cleaning or decontamination method except
those recommended by the manufacturer, users should check that the proposed
method would not damage the equipment.
8.2
Replacement of Fuses
There are two supply fuses located in the fuse drawer. To change the fuses:
• Turn power switch to the off position
• Disconnect the unit from the power supply
• Remove the line cord from the power entry module on the back of the unit
• Pull back on the fuse drawer (see sketch below)
• Pull out the fuse drawer
• Check and replace with the correct fuses if necessary. The fuses should be
5mm x 20mm quick acting, rated 250V.
Model 133000, 133000S (115 V AC):
1AF
Model 133000-2 (230 V AC):
.5AF
• Push the fuse drawer back in. Reconnect unit to power supply.