6. Fault Diagnosis


Symptom

Possible Cause

Action Required

1. Unit does not operate

a. Unit not switched on
b. Unit not plugged into

power supply

c. Fuses blown
d. Power supply failure

a. Switch on
b. Plug in, switch on

c. Replace fuses per 8.2
d. Check that other electrical

appliances on the same
circuit are working

2. Chamber temperature does

not rise when expected

a. Actual temperature is

higher than set
temperature

b. Temperature control

circuit fault

a. Check set temperature


b. Have unit checked by

competent person

3. Temperature continues to

rise when not expected

a. Actual temperature is

lower than set temperature

b. Temperature control

circuit fault

a. Check set temperature

b. Have unit checked by

competent person

4. Slides dry out during

heating cycle

a. Gasket not sealing

b. Lid bent
c. Insufficient buffer placed

in bottom of tray

d. Latch not fully engaged

a. Replace Tray and Gasket

Assembly

b. Replace Lid
c. Place more buffer in

bottom of tray

d. Re-install Tray Assembly

into chamber and re-
tighten Latch

7. Technical Specifications


This equipment is intended for indoor use and will meet its performance figures within the
ambient temperature range of 10

°C to 35°C, with maximum relative humidity of 80% (non-

condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance
with IEC 664. Suitable for operation at altitudes of up to 6500 feet.