6. Fault Diagnosis
Symptom
Possible Cause
Action Required
1. Unit does not operate
a. Unit not switched on
a. Switch on
b. Unit not plugged into
power supply
b. Plug in, switch on
c. Fuses blown
c. Replace fuses per 8.2
d. Power supply failure
d. Check that other
electrical appliances on
the same circuit are
working
2. Chamber temperature
does not rise when
expected
a. Actual temperature is
higher than set
temperature
a. Check set temperature
b. Temperature control
circuit fault
b. Have unit checked by
competent person
c. Timer cycle has ended
c. Turn timer off.
3. Temperature continues
to rise when not
expected
a. Actual temperature is
lower than set
temperature
a. Check setting
b. Temperature control
circuit fault
b. Have unit checked by
competent person
4. No rotation, Display
illuminated
a. Carousel switch not on
a. Turn motor jog switch on
b. An obstruction prevents
Carousel rotation
b. Remove obstruction
c. Carousel/Spindles are
loose
c. Tighten set screws w/hex
key