Maintenance and service, 1 cleaning, 2 replacement of fuse (see figure 1) – Boekel Scientific 177001 Aspirator Pump User Manual
Page 7: Warranty, Service

7
8.
Maintenance and Service
All Boekel laboratory products are designed to comply with
IEC1010-1. No routine maintenance is required.
8.1 Cleaning
Disengage power cord prior to cleaning. The aspirators should be
rinsed with fresh water after each use. The exterior of the unit can
be cleaned with a cloth dampened with water.
8.2 Replacement of Fuse (see figure 1)
To change the fuse:
Turn power switch to the off position. Disconnect the unit from the
power supply. Turn the fuse carrier in a counter clockwise direction
to loosen it. Remove the fuse from the fuse carrier. Check and
replace with the correct fuse if necessary. The fuse should be ¼” x 1
¼” Fast Blow, rated 5 AF, 250V. Push the fuse into the fuse carrier
and then turn the fuse carrier in a clockwise direction to tighten.
Reconnect unit to power supply. Replacement fuses (part number
902-0179) may be purchased from Boekel.
9.
Warranty
When used in laboratory conditions and according to these
operating instructions, Boekel warrants this product to be free of
defective material and workmanship for a period of two years from
the date of manufacture. The liability of Boekel for any defective
equipment during the warranty period shall be limited to the repair
of such equipment or replacement thereof without charge for parts
or labor.
10. Service
It is required to obtain a Returned Material Authorization (RMA)
number before any Boekel products are returned for any reason. A
Decontamination Certificate must be completed, signed by the
user, and returned to Boekel Scientific prior to receiving the RMA
number. Please be sure to mark the outside of the returned goods
package with this RMA number to ensure prompt handling.