Technical specifications, Accessories, Fault diagnosis – Boekel Scientific 177001 Aspirator Pump User Manual

Page 6

background image

6

5.

Technical Specifications

This equipment is for indoor use and will meet its performance

figures within the ambient temperature range of 10°C to 35°C, with

maximum relative humidity of 80% non-condensing. Installation

category II (transient voltages). Pollution degree 2 in accordance

with IEC 664. For operation at altitudes of up to 6500 feet.

Maximum Vacuum

29 “ of mercury

Maximum Suction Rate

14 liters per minute per port

Supply Ratings

Model 177001: 115 VAC, 2.7 A, 60 Hz

Power Rating

Model 177001: 310 W

Weight

6 Kg

6.

Accessories

Model Number

Description

177010

Pump Dolly

177020

Vacuum Regulator Gauge

7.

Fault Diagnosis

Symptom

Possible Cause

Action Required

Unit does not operate

a. Unit not switched on

b. Unit not plugged

into power supply

c. Fuse blown

d. Power supply failure

a. Switch on

b. Plug in, switch on

c. Replace fuse (see

section 8.2)

d. Check that other

electrical

appliances on the

same circuit are

working

Insufficient vacuum

a. Leak in system

b. Unit is

contaminated

c. Float ball stuck in

aspirator

a. Check all hoses

and hose

connections

b. Thoroughly clean

entire system

c. Tap lightly on the

aspirator or replace

aspirator.