Accessories, Block –14 – 1.5 ml tubes, Block – 14 x 0.5 ml tubes – Boekel Scientific 260013-2 Polar Block User Manual
Page 6: Block – 7 x 1.5 ml tubes plus, Well lid, Fault diagnosis, Technical specifications

5
5.
Accessories
5.1
Block –14 – 1.5 mL tubes
Part Number 260152
5.2
Block – 14 x 0.5 mL tubes
Part Number 260052
5.3
Block – 7 x 1.5 mL tubes plus
7 x 0.5 mL Tubes
Part Number 260055
5.4
Well Lid
Part Number A1702558
6. Fault Diagnosis
Symptom
Possible Cause
Action Required
1. Unit does not operate
a. Unit not switched on
b. Unit not plugged into
power supply
c. Fuses blown
d. Power supply failure
a. Switch on
b. Plug in, switch on
c. Replace fuses per 8.2
d. Check that other electrical
appliances on the same
circuit are working
2. Well temperature does not
cool when expected
a. Temperature control
circuit fault
a. Have unit checked by
competent person
3. Well temperature continues
to cool when not expected
a. Temperature control
circuit fault
a. Have unit checked by
competent person
7. Technical Specifications
This equipment is intended for indoor use and will meet its performance figures within
the ambient temperature range of 10
°C to 35°C, with maximum relative humidity of 80%
(non-condensing). Installation Category II (transient voltages). Pollution Degree 2 in
accordance with IEC 664. Suitable for operation at altitudes of up to 6500 feet.