Technical specifications, Maintenance and service, 1 cleaning – Boekel Scientific 260014-2 TropiCooler User Manual
Page 7: 2 replacement of fuses

6
b. Temperature control
circuit fault
b. Have unit checked by
competent person
4. Well temperature does not
cool when not expected
a. Actual temperature is
lower than set temperature
b. Temperature control
circuit fault
a. Check set temperature
b. Have unit checked by
competent person
5. Well temperature continues
to cool when not expected
a. Actual temperature is
higher than set
temperature
b. Temperature control
circuit fault
a. Check set temperature
b. Have unit checked by
competent person
7. Technical Specifications
This equipment is intended for indoor use and will meet its performance figures within the
ambient temperature range of 10
°C to 35°C, with maximum relative humidity of 80% (non-
condensing). Installation Category II (transient voltages). Pollution Degree 2 in accordance
with IEC 664. Suitable for operation at altitudes of up to 6500 feet.
Specifications:
Temperature Range:
39
°C below ambient to 69°C
Supply Voltage Range:
115V +/- 10%, 50/60 Hz
230V +/- 10%, 50/60 Hz
Power Rating:
Model 260014: 75W, 1.0A
Model 260014-2: 75W, 0.5A
8. Maintenance and Service
All Boekel laboratory products are designed to comply with IEC1010-1. It is
recommended that the unit be turned off and defrosted if substantial frost builds up on the
block or in the well. This will allow the TropiCooler to operate at maximum efficiency.
8.1
Cleaning
Disengage power cord prior to cleaning. The outer casing may be cleaned with water and
a damp cloth. Do not submerge or immerse the TropiCooler in water. Abrasive
materials such as paper towels may scratch the well cover or the plastic body. Do not use
solvents to clean the well cover or the plastic body. Before using any cleaning or
decontamination method, users should check that the proposed method would not damage
the equipment.
8.2
Replacement of Fuses
There are two supply fuses located in the fuse drawer. To replace the fuses: