Specifications, Standards – American Diagnostic Corporation (ADC) e-sphyg 7002 User Manual
Page 13

Specifications
Power Source: Two 1.5V DC LR03 (AAA) Alkaline Batteries
MEASUREMENT RANGE
Pressure: 0-299 mmHg
Pulse: 30-240 beats/min Accuracy
Pressure: ±3 mmHg
Pulse: Within ±5% of reading
Pressure Sensor: Semi-conductor
Inflation: Manual
Deflation: Manual Air Release Valve
Auto-shut-off: 1 min. after last key operation
Backlight auto-shut-off: 15 sec. after last key operation
OPERATION ENVIRONMENT
Temperature: 32~115°F (0~46°C)
Humidity: 85% Relative Humidity max
STORAGE ENVIRONMENT
Temperature: -4~131°F (-20~55°C)
Humidity: 90% Relative Humidity max (non condensing)
Dimensions: 3”(L) X 2
3/8
”(W) X 1
1/8
”(H)
Weight: 3.14 oz. (G.W.) (w/o Batteries) Type BF
Battery Life: 1500 readings with back light on
3000 readings with back light off
Device and cuff are designed to provide special protection against electrical shocks.
Specifications are subject to change without notice.
Standards
ADC’s digital aneroid sphygmomanometer is accurate to +/- 3 mmHg
and conforms to applicable sections of the following international standards:
EN /ISO 81060-1: 2012, Non-invasive sphygmomanometers - Part 1:
General Requirements
EN /ISO 81060-1: 2012, Non-invasive sphygmomanometers - Part 3:
supplementary requirements for electro-mechanical blood pressure measuring systems
ANSI/AAMI/ISO 81060-1: 2007, Manual, Electronic, or Automated Sphygmomanometers
Symbol
Definition
Important Warning/Caution
Latex-Free
Circumference Size
Conforms to EU Standards
Symbol
Definition
Authorized European
Represenative’s Information
Manufacturer’s Information
Temperature Limit
Humidity Limitation
°
F