beautypg.com

3 releasing the torquemaster, 4 torquemaster malfunctions, 5 compatibility – FRITSCH PULVERISETTE 0 User Manual

Page 37: Accessories

background image

7.2.7.3

Releasing the TorqueMaster

The TorqueMaster clamping device is also released with the cordless

screwdriver.
The clamping unit is released far enough so that it can be placed behind

the device without changing the toothed belt clamp in the belt clamping

device.
The clamped toothed belts on the belt clamping device is only released

or adjusted, when the number or height of the sieves in the sieve stack

changes.

7.2.7.4

TorqueMaster malfunctions

For all functions, the two indicators of the clamping unit must be in the

green area. If one or both indicators are in the red area, proceed

according to the following table.

Malfunction

Possible causes

Rectify fault

Both indicators in outer red area

Clamping unit open too much

Open the clamping unit to a lesser

degree, the toothed belt clamp is

possibly too tight and must be loos-

ened by one tooth at the belt

clamping device on both sides.

Clamping unit must be opened far

enough to allow it to be deposited /

clamping unit not attached according

to instructions

Attach according to instructions.

Both indicators in inner red area

respectively

Clamping unit closed too much /

clamping unit not attached according

to instructions

Attach according to instructions.

An indicator in red area on one side

Clamping system attached "very"

asymmetrically

Attach according to instructions.

Make sure that the toothed belt is

fastened symmetrically in the belt

clamping device.

7.2.7.5

Compatibility

The TorqueMaster must be used with the toothed belt clamping device

(03.3120.00). The design and proper use of this combination will rule out

any damage to the toothed belt.
Generally, the TorqueMaster can also be used with the older belt

clamping device (03.1210.00). However, this combination leads to

increased wear and is therefore not recommended.

Accessories