miniQV-X Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.0— October, 2008; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
48
Use the ID field to input instrument identification information using up to
16 alphanumeric characters.
The S/N: field displays the instrument serial number stored by the
operational firmware of the miniQV-X. This is a non-editable field.
The miniQV-X Instrument Settings window
Click on this check box to activate a software counter which will gener-
ate a warning for a tube whenever the specified number of days have
elapsed without performance of the Verify Known KV sample action.
Enter the number of days in the appropriate field.
Click on this check box to activate a software counter which will gener-
ate a warning whenever the specified number of samples have been
tested without performance of the Verify Known KV sample action.
Enter the number of samples in the appropriate field.