BUCHI SpeedDigester K-439 User Manual
Page 32

6 Operation
32
K-439 Operation Manual, Version A
Note
If the rack(s) is/are cooled down inside the heating chamber, the time for the cool down step in the
corresponding method must be set to
≥ 80 min, so that no harmful vapors can escape. If the racks
are cooled outside the heating chamber, the cool down step in the corresponding method must be
set to
≥ 30 min.
Press Save.
•
Now enter a name for the new method by highlighting individual letters on the virtual keyboard with
•
the ± buttons and press Enter to confirm a letter.
Repeat this procedure until the name is completed.
•
Press Save.
•
The new method now appears in the method list and can be loaded to start a digestion.