beautypg.com

8 troubleshooting, 1 malfunctions and their remedy, Malfunctions and their remedy – BUCHI Melting Point M-560 User Manual

Page 41: 8troubleshooting

background image

8 Troubleshooting

41

Melting Point M-560 Operation Manual, Version D

8

Troubleshooting

The following section describes how to resume operation of the instrument in the event of any minor
problem. It lists some possible occurrences, their probable cause, and suggests how to remedy
the problem. The troubleshooting table below lists possible instrument malfunctions and errors and
describes courses of action that can be used by operators to correct some of those problems. The
appropriate course of action is listed in the column “Remedy”.
More complicated malfunctions or errors are usually handled by a BUCHI technical engineer who has
access to the official service manuals. In this case, please contact your local BUCHI customer service
agent.

8.1

Malfunctions and their remedy

Table 8-1: General malfunctions and their remedy

Malfunction

Possible cause

Remedy

Instrument does not work

Main switch off

Switch on mains switch

Instrument is not connected to mains
supply

Check mains connection

No or unreadable printout

Not activated in the settings

Activate automatic printout in the
settings

Bad cable connection

Check cable connection

Printer switched OFF

Switch ON printer

No paper

Replace paper roll

Baud rate settings of printer faulty

See section 8.1.1

No or wrong result

Sample-specific preparations:
Sample is not dry or is contaminated
by another substance

Sample should be dried before use

Sample decomposes during the
melting process (formation of
bubbles, sample turns brown, etc.)
Apparatus parameters:
Apparatus is not or is poorly cali-
brated

Regular calibration of apparatus with
BUCHI calibration standards or check
with verification standards

Setting point is too close to the
melting point

Select setting point 5–10 °C below
melting point

Instrument does not heat

Heating defective

Contact BUCHI customer service.

Table 8-2: Warning messages

Warning number

Possible cause

Remedy

Warning 01

Calibration outdated.

Calibrate or verify device.

Warning 02

Housing temperature too high.

Cool down device and check clearance of vent
holes and fans.
Check ambient temperature.

Warning 03

Device not calibrated.

Calibrate device.