3 Technical data
12
Melting Point M-560 Operation Manual, Version D
Table 3-2: Optional accessories (cont.)
Product
Order number
Ribbon for printer
044306
Paper roll for printer
038684
A) Boiling point tubes, 100 units
019007
B) Boiling point capillaries, 100 units
051890
Protective cover
051935
3.2
Technical data overview
Table 3-4: Technical data of the Melting Point M-560
Melting Point M-560
Manual melting point determination
√
Manual boiling point determination
√
Automatic melting point determination
–
Automatic boiling point determination
–
Homogeneous sample loading
–
Positions for melting capillaries
3
Positions for boiling capillaries
1
Precision magnifying lens
√
Magnification of lens
2.5×
Display
Colour, TFT, 320×240, 3.5˝
Determination temperature range
Ambient +10 °C to 400 °C
Temperature resolution
0.1 °C
Accuracy of melting point at 0.5 °C/min
up to 250 °C
±0.3 °C