8 Troubleshooting

47

V-700/710 Operation Manual, Version E

8 Troubleshooting

This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a BUCHI technical
engineer who has access to the official service manuals. In this case, please refer to your local BUCHI
customer service agent.

8 .1

Malfunctions and their remedy

Table 8-1: General malfunction and their remedy

Malfunction

Possible cause

Corrective measure

Instrument does not work

Mains switch off

Switch on mains switch

Instrument is not connected to power
supply

Check if mains connection is okay

Instrument is not connected to power
supply

Check mains connection

System is leaky or does not reach
the end vacuum

Tube clips have not been fixed
correctly or are defective

Check tube clips

Tubes are leaky (brittle)

Replace tubes

Membrane and/or valves are
contaminated

Clean or replace membrane and/or
valves

Contact at the removable top cover is
not closed

Top cover not perfectly mounted

Check the mounting of the top cover
and refix it, if necessary