2 Safety

9

R II Operation Manual, Version E

WARNING
Potential explosion risk if solvent vapors accumulate within the instrument housing.
• Always use the instrument in a well ventilated area.
• Beware of damaged or cracked glass parts.
• Beware of the fire hazard.

2 .5 .2

Other hazards

WARNING
Certain solvents in or in the vicinity of the Rotavapor can form peroxides and/or are highly inflam-
mable.
• Always be aware of the explosion risk when working with hazardous substances or with

substances of unknown composition.

• Always provide a good ventilation within or in the vicinity of the system.

2 .5 .3

Safety measures

Always wear personal protective equipment such as protective eye goggles, protective clothing and
gloves when working with the instrument.

2 .5 .4

Safety elements

Electronics
• The heating bath is equipped with a mechanical over-temperature protection.

The mechanical over-temperature protection consists of a bimetal thermostat that, in case of over-
temperature (over 260 °C), directly interrupts the power supply. It has to be set back manually
after the bath has cooled down (see also chapter 8).
The electronic over-temperature protection controls the temperature limit (actual bath temperature
may not exceed the given temperature by 2 °C for more than 2 minutes), the heating rate (actual
temperature may not rise by more than 5 °C during 5 seconds) and the function of the tempera-
ture sensor.

• The heating bath is equipped with safety fuses.
• The heating bath is equipped with a thermostatic control of the bath temperature - to prevent the

product from overheating

Parts in direct contact with the instrument
• Combi-Clips for fixing the evaporating flask and for safe removal of fixed ground-glass joints.
• Ball joint clip for safe fixing of the receiving flask.
• Rods and holders for fixing the glass assemblies.
• Electronic over-current protection at the drive unit.
• Safety spring preventing the vapor duct from dropping out.

Glass
• Use of high quality, inert 3.3 borosilicate glass.
• Use of tube connections GL 14 for preventing glass breakage.