beautypg.com

Fat extraction calculations, Fat extraction support items, Operator’s manual – ANKOM XT10 Series User Manual

Page 10

background image

Operator’s Manual


pg. 10

Rev F 12/08/14

Fat Extraction Calculations

Crude Fat Calculation

Crude Fat contained within a food or feed sample can be calculated using the following formula:

% Crude Fat

=

100 x (W

2

– W

3

)

W

1

Where:

W

1

=

Original weight of sample

W

2

=

Weight of pre-extraction dried sample and filter bag

W

3

=

Weight of dried sample and filter bag after extraction

For Liquid Samples,

W

2

and

W

3

in the equation above will include the weight of Diatomaceous Earth (DE).

For meat samples with > 15% fat and plant samples with > 20% fat:

% Crude Fat

=

100 x ((W

2

– W

4

) – W

3

)

W

1

Where:

W

4

=

Weight of the weigh tin

Total Fat Calculation

If hydrolysis is done prior to fat extraction, then Total Fat contained within a food or feed sample can be calculated
using the following formula:

% Total Fat

=

100 x (W

2

– (W

3

+ (C

1

– C

2

)))

W

1

Where:

W

1

=

Original weight of sample

W

2

=

Weight of dried sample, filter bag, and Diatomaceous Earth (DE) after hydrolysis

W

3

=

Weight of dry extracted sample, filter bag, and DE

C

1

=

Blank filter bag dry weight after hydrolysis

C

2

=

Blank filter bag weight after extraction

Fat Extraction Support Items

The following support items are needed to perform the fat extraction procedure:

Item

Recommended Product

Electronic Balance with four-place readout

ANKOM #TB Balance Hardware
ANKOM #TBS Balance Software

Filter Bags

ANKOM #XT4

Bag Holder (used for adding sample to an empty filter bag)

ANKOM #101.2

Heat Sealer for sealing the filter bags

ANKOM #1915 (120V), #1920 (220V)

Solvent Resistant Marker

ANKOM #F08

Desiccant Pouch or Weigh Tin Desiccator for drying

ANKOM #X45, #X49

Dryer (capable of temperatures from 60-120°C) - optional

ANKOM #RD (120V), #RDI (220V)

Oven for drying (capable of maintaining 102°C±2°)

Ventilated oven trays to allow for sufficient oven air flow

Weigh Tins (used with samples of > 20% fat)

Sample

Spoon