Compur statox 4120, The maintenace request option – Compur STATOX 4120 User Manual
Page 25

Compur Statox 4120
 
 
 
 
 
7. The maintenace request option
 
Purpose: 
 
This option can differentiate between fatal errors in the sensor head and maintenance requirements. It 
requires a Statox 4120 control module equipped with EPROM Index ≥ 03. It also sets the analog 
output to 4 mA during the self test, if required. 
 
 
Operation modes: 
 
The Statox 4120 can operate in 9 different modes. The relevant mode can be seen by the control 
module LED`s and relays. 
 
 
 
Operation mode Statox
Analog output
SR-LED
Control 
module 
SF-LED
Control 
module 
SF-Relay MR-
Relay
PCS-*
Relay
O
Self test
0mA / 4mA / 0V
„on“
„on“
active
passive
active
O
Diagnostic box connected:
0mA / 4mA / 0V
flashing „off“ active
passive
active
O
Alarm threshold adjustment:
0mA / 4mA / 0V „on“
„off“ active passive
active
O
Start-mode
4mA / 0V
flashing
„off“
active
passive
passive
O
Detection mode:
Analog signal
„on“
„off“
active
passive
passive
O
Maintenance request:
B1 Battery Low 
B2 Battery low under load 
C1 Purge before test 
C2 Purge between test peaks 
C3 Sensitivity 
 (programmable by S3) 
C4 Purge 
PU Pump current too high 
T2 Temperature out of specs
 
 
 
 
 
 
Analog signal
 
 
 
 
 
 
flashing
 
 
 
 
 
 
„off“
 
 
 
 
 
 
active
 
 
 
 
 
 
active
 
 
 
 
 
 
passive
O
System failure:
A1 Amplifier 
C3 Sensitivity 
(programmable by S3) 
 
 
4mA / 0V
 
 
„off“
 
 
„on“
 
 
open
 
 
passive
 
 
passive
O
Data transmission error
4mA/0V
„off“
„on“
open
passive
passive
O
Power failure
0mA/0V
„off“ „off“
open
passive
passive
 
*(PCS=Process control system) 
 
 
 
 
 
 
 
 
25