Compur statox 4120, Trouble shooting, Sf flashing sf on – Compur STATOX 4120 User Manual
Page 19

Compur Statox 4120
 
 
 
 
 
4.5 Manually triggered self test of the system
 
Push and hold button T for 2 s to initiate a self test. 
The next self test will start automatically after 24 h. 
 
 
4.6 System failure
 
As soon as the Statox detects a faulty system condition the LED SF goes on and relay SF is activated. 
This will be the case if: 
 
Self test not passed
Data transmission faulty
No signal
5. Trouble shooting
 
As soon as the Statox detects a faulty system condition the LED SF goes on and relay SF is activated. 
To locate the problem follow this procedure: 
 
Push the reset button R on the control module. 
If the green LED starts flashing now, there is a data transmission problem.
If there is no reaction, there is a hardware problem in the sensor head (f. i. pump, sensor, gas
generator, battery).
Now push button T to initiate a new self test. If SF goes off, the system is now ready again. If SF
 stays on, the sensor head must be checked with the diagnostic box. 
 
R
Push button R
 
 
 
 
SF flashing
SF ON
 
 
 
 
 
 
 
Data transmission problem. 
Potential errors: 
Signal cable interrupted
Wrong connection(SF goes on again
after 4 min)
Data transmission faulty
Duty cycle needs adjustment
Connect diagnostic box to sensor 
head. Potential errors: 
 
Sensor defective
 Pump defective 
 Battery defective 
Amplifier defective
 
 
 
 
 
 
 
 
 
 
 
 
19