Table 2: switch position for hcn, Compur statox 4120 – Compur STATOX 4120 User Manual
Page 16

Compur Statox 4120
DIP switch S1 (see pic. 8) programs gas and measuring range. If you change your Statox to another
gas or measuring range, a hardware modification may be required and the electronics must be re -
adjusted. This must be done by authorised personnel.
Table 1: Switch position for H
2
S
Measuring range:
0-30 ppm
0-50 ppm
0-100 ppm
S1/1 ON
OFF
OFF
S1/2 ON
ON
ON
S1/3 ON
ON
ON
S1/4 ON
ON
OFF
S1/5 ON
ON
OFF
S1/6 ON
OFF
ON
S1/7 ON
ON
OFF
S1/8 ON
ON
ON
E-PROM Index sensor head
12, 13, 14
12, 13, 14
12, 13, 14
E-PROM Index control module
≥ 02
≥ 02
≥ 02
Table 2: Switch position for HCN
Measuring range:
0-15 ppm 0-30 ppm 0-50 ppm 0-100 ppm
S1/1 ON
ON
OFF
OFF
S1/2
ON ON ON ON
S1/3
ON ON ON ON
S1/4 OFF
OFF
OFF
ON
S1/5
OFF ON ON ON
S1/6
OFF ON OFF ON
S1/7
OFF ON ON ON
S1/8
ON ON ON ON
E-PROM Index sensor head
13
12, 13, 14
12, 13, 14
15
E-PROM Index control module
≥ 02
≥ 02
≥ 02
≥ 02
Table 3: Switch position for COCl
2
Measuring range:
0-0,1 ppm
0-0,3 ppm
0-0,3 ppm 0-0,5 ppm 0-1,5 ppm
0-15 ppm
0-100 ppm
S1/1
OFF ON ON OFF OFF OFF OFF
S1/2
ON ON ON ON ON ON ON
S1/3
ON ON ON ON ON ON ON
S1/4 OFF
ON
ON
ON OFF ON ON
S1/5
ON OFF ON OFF ON OFF OFF
S1/6 ON
ON
ON
OFF OFF OFF ON
S1/7 ON
ON
ON
ON OFF OFF ON
S1/8
ON ON ON ON ON ON ON
E-PROM Index
sensor head
20
12, 13, 14
18
12, 13, 14
12, 13, 14
12, 13, 14
17
E-PROM Index
control module
04
≥
02
≥
02
≥
02
≥
02
≥
02
≥
02
16