beautypg.com

Pilgrim Model K1 Pressure Transmitter User Manual

Model k1 pressure transmitter

background image

Model K1 Pressure Transmitter

BULLETIN PT-1

APPLICATIONS

Hydraulic, refrigeration, machine tool,
test/measurement, pump control, HVAC,
medical, construction equipment and
all general purpose industrial process
applications

FEATURES:

• 0.5% and 1.0% accuracy

• Vac.-20,000 psi pressure range

• FM approved (optional)

• Superior long-term stability and

repeatability

• Stainless steel NEMA 4X enclosure

• Current/voltage output

• Wide range of electrical connections

available

• Optional panel meter digital display –

see Ashcroft model 2269

The Ashcroft

®

K1 is a proven and versa-

tile pressure transducer/transmitter incor-
porating polysilicon thin film technology.
Modern low-pressure chemical vapor dep-
osition methods provide simple, stable
molecular bonds between the metal
diaphragm and a polysilicon strain gage
bridge. There are no epoxies or bonding
agents to contribute to signal instability
or drift. The integral metal diaphragm and
polysilicon bridge are virtually unaffected
by shock, vibration or mounting.

These transmitters are offered in many

standard pressure ranges with either cur-
rent or voltage output signals. Transmitter
performance is directly traceable to the
NationaI Institute of Standards and
Technology. A calibration test certificate is
available with each transmitter.

PERFORMANCE SPECIFICATIONS

Accuracy Class (Span):

±0.5%

±1.0%

Includes non-linearity
(Terminal Point Method), hystere-
sis, non-repeatability, zero offset
and span setting errors

Best Fit Straight Line (BFSL)

±0.25%

±0.4%

Non-linearity

Interchangeability

±0.5%

±1.0%

Durability: 10

8

cycles 20/80% span with negligible

performance change

Stability: ±0.5% Span/yr

ENVIRONMENTAL SPECIFICATIONS

Temperature Limits:
Storage:

–54 to 121°C

(–65 to +250°F)

Operating:

–28 to 82°C

(–20 to +180°F)

Comp. Range:

–28 to 71°C

(–20 to +160°F)

Thermal Coefficients: (68°F (20°C) ref.) % Span/°F:
Accuracy Class: 0.5%

1%

ZERO

±0.028%

±0.04%

SPAN

±0.028%

±0.04%

Optional (0.5% Accuracy Class only):
ZERO

±0.014%

N/A

SPAN

±0.014%

N/A

Multiply thermal zero coefficients by 1.5 on 0/30
psi, vac/15 range and by 3 on 0/15 and vac/0
ranges

Humidity: No performance effect at 95% relative
humidity-noncondensing

FUNCTIONAL SPECIFICATIONS

Standard Ranges (psi) gauge, compound:
0/15*

0/300

0/5000

vac./60*

0/30*

0/500

0/7500*

vac./45*

0/60*

0/750

0/10,000* vac./30*

0/100

0/1000

0/15,000* vac./15*

0/150

0/2000

0/20,000* vac./0*

0/200

0/3000

*1% accuracy ranges only.

Consult factory for nonstandard ranges\.

Overpressure Limits (F.S.):

15-

3000-

7500-

2000

5000

20,000

Proof

200%

150%

120%

Burst

800%

300%

150%

Vibration Sweep: Less than ±0.1%F.S. effect for
0-2000 Hz at 20 g’s in any axis

Shock: Less than ±0.05% F.S. effect for 100 g's,
20ms shock in any axis

Response Time: Less than 5 ms

Position Effect: Less than 0.01% F.S.

ELECTRICAL SPECIFICATIONS

Output Signal (consult factory for options):
4-20mA (2 wire)
1-5 Vdc (3 wire)
1-6 Vdc (3 wire)
1-11 Vdc (3 wire) (minimum excitation 15 Vdc)

Power Requirements: 10-36 Vdc unregulated,
<3mA for voltage output

LOOK FOR THIS AGENCY

MARK ON OUR PRODUCTS

Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
Tel: 203-378-8281 • Fax: 203-385-0408
email: [email protected] • www.ashcroft.com

All specifications are subject to change without notice.
All sales subject to standard terms and conditions.
© Ashcroft Inc. 2009 11/10

Ash Bul PT-1:Layout

11/12/10

1:59 PM

Page 1