Vetter Docking tent User Manual
Page 4

eg. Tent blower 0.23-0.47 bar
(Art. no. 1523017700)
Vetter Medical Tents
Page 3/12
9
Screening facility for treatment of injuries
9
Accommodation tent
9
Emergency accommodation during evacuations
The medical tents may only be operated with compressed air;
never use flammable or poisonous gases. The medical tents may
only be filled using original Vetter fixtures.
2.4 Safety instructions
Only trained, reliable, and responsible employees are permitted
to use and operate this product. Wear the required protective
gear and clothing. Please contact the manufacturer or authorized
retailer when detecting cracks, fissures, holes, or leakages on the
pneumatic supporting frame during the product inspection. Such
defects can lead to bursting. Always avoid contact with hot or
sharp-edged materials.
Always fix tents in place using tension ropes
and, depending on the wind load, use
additional ballast tanks (filled with water)!
3.
Preparing the product for use
3.1 Preparations for operation
Inspect the medical tents before and after each operation to en-
sure correct and perfect condition. It is imperative that the recom-
mended folding instructions be observed.
4.
Operating Instructions
Open the packing cover. Lay out the tent at the required location.
The ground area must be free of any sharp-edged or pointed ob-
jects!
The pneumatic medical tents are designed to be inflated with a
blower or with compressed air.
4.1 Operation with blower
The inflation connection for blowers is opened when air flows
through. Connect the inflation hose of the blower to the inflation
connection of the tent. Unscrew the cap from the inflation connec-
tion in a counter-clockwise direction and check whether the valve
is closed. Make certain there is perfect locking of the couplings of
the inflation hose with the tent coupling and the blower.
Start the blower and fill the tent until the maximum permitted
operating pressure of 0.3 bar/4.35 psi. The medical tent erects
automatically. The safety valve integrated into the support frame
avoids over-inflation (greater than 0.3 bar/4.35 psi).