Guidance and manufacture’s declaration – electromagnetic immunity –

for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7HC should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or

discharge (ESD)

±8 kV air

±8 kV air

ceramic tile. If floor are covered with

IEC 61000-4-2

synthetic material, the relative humidity

should be at least 30%.

Electrical fast

±2 kV for power

±2kV for power

Mains power quality should be that of

transient/burst

supply lines

supply lines

a typical commercial or hospital IEC

61000-4-4

environment.

Surge

± 1 kV line(s) t

±1 kV differential

Mains power quality should be that of

IEC 61000-4-5

o line(s)

mode

a typical commercial or hospital

environment.

Voltage dips, short

<5% U

T

<5% U

T

Mains power quality should be that

interruptions and

(>95% dip in U

T

)

(>95% dip in U

T

)

of a typical commercial or hospital

voltage variations

for 0.5 cycle

for 0.5 cycle

environment. If the user of the Pro7HC

on power supply

requires continued operation during

input lines

40% U

T

40% U

T

power mains interruptions, it is IEC

61000-4-11

(60% dip in U

T

)

(60% dip in U

T

)

recommended that the Pro7HC be

powered

for 5 cycles

for 5 cycles from an

uninterruptible power supply or a

battery.

70% U

T

70% U

T

(30% dip in U

T

)

(30% dip in U

T

)

for 25 cycles

for 25 cycles

<5% U

T

<5% U

T

(>95% dip in U

T

)

(>95% dip in U

T

)

for 5 sec

for 5 sec

Power frequency

3A/m

3A/m

Power frequency magnetic fields should

(50Hz) magnetic

be at levels characteristic of a typical

61000-4-8

location in a typical commercial or

hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.

13

Guidance and manufacture’s declaration – electromagnetic immunity –

for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7HC should assure that it is used in such an environment.

Immunity test

IEC 60601 test

Compliance

Electromagnetic environment

level

level -

guidance

Portable and mobile RF communications

equipment should be used no closer to

any part of the Pro7HC, including cables, than

the recommended separation distance

calculated from the equation applicable to the

frequency of the transmitter.

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

IEC 61000-4-6

150 kHz to 80 MHz

Radiated RF

3 V/m

3 Vrms

80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

800 MHz to 2.5 GHz

Where

P is the maximum output power rating

of the transmitter in watts (W) according

to the transmitter manufacturer and d is the

recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in each

frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following symbol:



NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Pro7HC is used exceeds
the applicable RF compliance level above, the Pro7HC should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro7HC.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

P

d

167

.

1

=

P

d

167

.

1

=

P

d

167

.

1

=

14