Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7HC should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood, concrete or
discharge (ESD)
±8 kV air
±8 kV air
ceramic tile. If floor are covered with
IEC 61000-4-2
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
±2 kV for power
±2kV for power
Mains power quality should be that of
transient/burst
supply lines
supply lines
a typical commercial or hospital IEC
61000-4-4
environment.
Surge
± 1 kV line(s) t
±1 kV differential
Mains power quality should be that of
IEC 61000-4-5
o line(s)
mode
a typical commercial or hospital
environment.
Voltage dips, short
<5% U
T
<5% U
T
Mains power quality should be that
interruptions and
(>95% dip in U
T
)
(>95% dip in U
T
)
of a typical commercial or hospital
voltage variations
for 0.5 cycle
for 0.5 cycle
environment. If the user of the Pro7HC
on power supply
requires continued operation during
input lines
40% U
T
40% U
T
power mains interruptions, it is IEC
61000-4-11
(60% dip in U
T
)
(60% dip in U
T
)
recommended that the Pro7HC be
powered
for 5 cycles
for 5 cycles from an
uninterruptible power supply or a
battery.
70% U
T
70% U
T
(30% dip in U
T
)
(30% dip in U
T
)
for 25 cycles
for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
)
(>95% dip in U
T
)
for 5 sec
for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic fields should
(50Hz) magnetic
be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
13
Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7HC is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7HC should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Compliance
Electromagnetic environment
level
level -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Pro7HC, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
3 Vrms
80 MHz to 800 MHz
IEC 61000-4-3
80 MHz to 2.5 GHz
800 MHz to 2.5 GHz
Where
P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Pro7HC is used exceeds
the applicable RF compliance level above, the Pro7HC should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro7HC.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
d
167
.
1
=
P
d
167
.
1
=
P
d
167
.
1
=
14