beautypg.com

Type a standard, Laboratory airflow monitors & controls, Schematic aircontrol a led & digital – Plastec Airflow Monitor A User Manual

Page 4: Led & digital

background image

115 VAC / 12 VDC
Power Supply

Controller

Sensor

Sensor

Signal Input

Sash High Alarm Input

Sash movement

=

Pressure decrease

or increase

Hood Light
on/off relay

Air Make-up

on /off relay

Motorized

Damper

OR

Make-up

Fan

PLASTEC

®

Extraction Fan

Air speed

monitoring & control

!

If air speed is low

OUTPUT:

0 to 10V

or

4-20mA

Motor

Starter

Relay

Fan on /off

relay

2216 60th Drive East • Bradenton, Florida 34203

(941) 751-7596 • Fax (941) 751-7598 •

www.plastecvent.net

4

Schematic AirControl A LED & Digital

Laboratory Airflow Monitors & Controls

TYPE A STANDARD

LED & DIGITAL