MAINTENANCE
Front plate and stickers of airflow monitor may be cleaned with mild soap and water on a damp
cloth to remove finger marks, oils and residue. Do not use abrasives. Do not allow liquids to enter
the plastic casing. Dry the monitor thoroughly after cleaning.
TROUBLESHOOTING
PROBLEM
CHECK
WARRANTY
PLASTEC
®
Ventilation, Inc. warrants its equipment, products and parts, to be free from defects in
workmanship and material under normal use and service for one year after delivery to the first
user. Product must be returned to point of purchase, with dated bill of sale, within one year of
purchase. If factory return is required, please contact distributor first.
No indicators
Power supply may not be plugged into AC supply.
Improper alarm setpoint
Airflow sensor is factory precalibrated at 0.5 m/s or 100 FPM.
Field reading may be different if sensor is not properly located
or hood design affects stable airflow. If intended air speed is
different than 0.5 m/s / 100 FPM or if actual anemometer reading
is different, follow Controller Set Up Procedure on page 6.