beautypg.com

ETS-Lindgren HI-3627 ELF Magnetic Field Meter User Manual

Page 29

background image

HI-3627 Manual

Page

--21

7.0

Principle of Operation

The circuitry of the HI-3627 can be divided into six sections;

the probe, input signal conditioning, filter section, RMS

calculation, information processing and power supply. Each of

these is discussed in the follow sections (see Figure 7-1).

Probe

The probe consists of three concentric shielded coils aligned

with their axes perpendicular to one another. The electrostatic

shield of each coil is isolated from that of the other. The shields

are connected together at a single point at the base of the

sphere defined by the coils. The signals from each coil are

brought to the meter through a shielded twisted pair cable with

its shield connected to the coil shield common point.

Input Signal Conditioning

The output voltage of a coil varies with the frequency of the

applied magnetic field. A single pole low pass filter is added to

the output of each coil to flatten the frequency response and

provide impedance matching. The input signals are then passed

through a programmable gain amplifier stage that provides the

necessary amplification for the individual ranges. The signals

then pass to the filter section.

Filter Section

Each channel is filtered separately. The first stage is a two pole

Butterworth low pass filter at 2000 Hz. A Butterworth filter

attenuates the signal by 3 dB at the cutoff frequency with

attenuation in the stop band and no ripple in the pass band. The

output of the low pass filter is fed into two 4 pole high pass

filters. The cutoff frequencies on these filters are 5 Hz and 30

Hz corresponding to the low frequency cutoff selections.