ETS-Lindgren HI-3627 ELF Magnetic Field Meter User Manual

Page 12

background image

Page -- 4

HI-3627 Manual

Sensitivity:

Full scale ranges of:

2 mG

20 mG

200 mG

2 G

20 G

Accuracy:

± 5 % at calibration frequency

Power:

One 9.6 volt NiCad battery

Battery life:

Approximately 30 Hrs. continuous operation

(new battery with a full charge)

Linearity:

± 2% Full Scale

The instrument is calibrated using a one meter diameter pair of

Helmholtz coils for establishing an accurately known magnetic

field flux density. A precisely controlled and measured sinusoidal

current is driven through the Helmholtz coils and, based on the

dimensions of the coils, the magnetic field flux density between

the coils is calculated.

While the HI-3627 indicates magnetic flux density (B) in units

of milligauss, the flux density in microtesla or magnetic field

strength (H) in milliamperes per meter may be obtained via the

following relations:

1 mG = .1 microtesla (

:

T)

1 mG = 80 milliamperes per meter (mA/m)